Intended Use
The BioAid® Lipid Control Solution is intended for validating lipid testing while using the BioAid® Lipid Testing System. lipid control Solution contains stabilizers, preservatives and added chemicals. High-density lipoprotein (HDL) and triglyceride (TG) areincluded in the same control solution. Total Cholesterol (TC) is an individual control solution. lipid control Solution is applied intothe specimen well of the BioAid® Lipid Test Device which has been already inserted in the meter. The onscreen result confirmsthat the test device and meter are working together properly and that the test is being performed correctly. The control solution is available in two levels and is ready for validating the testing of TC, HDL and TG. It is for professional use.
● Use before the expiration date printed on the bottle label.
● Make sure the control solution and all the test materials reach operating temperature of 20-40℃ prior to testing. The control solutions and test materials are only accurate within this temperature range.
● All materials should be considered potentially hazardous and handled in the same manner as infectious agents.
● This product is not intended for use as a calibrator.
● Discard the control solution if it appears cloudy.
● Use the quality control materials before blood testing as an objective method to assess the techniques or practices in use.
● Use the BioAid® Lipid Control Solution with the BioAid® Lipid meter and test devices.
● The used device should be discarded according to local regulations after testing.
● Check the code chip before performing a test. Make sure to use the code chip that is included with the box of test devices.

Specification
|
Components |
No human resource materials |
|
Levels |
TC level 1 and level 2, HDL/TG level 1 and level 2 |
|
Amount |
2mL/bottle |
|
Shelf life |
12 months |
|
Open bottle shelf life |
3 months |
|
Store temperature |
2-30℃ |
|
Operation temperature |
20-40℃ |
Why to test
Lipid control solution is tested to ensure that the instrument used to measure lipid levels (like cholesterol and triglycerides) is working accurately. This quality control check is crucial because it confirms that the device provides reliable results, which is essential for correctly assessing a person's risk for cardiovascular disease and for monitoring the effectiveness of treatments.
Reasons for testing lipid control solution
● To ensure accuracy:
Lipid control solution is a reference material with known lipid values. By testing it, labs and doctors can verify that their testing equipment is providing correct measurements, just as a control solution is used for other diagnostic tests.
● To confirm proper function:
It ensures that a testing device, such as a point-of-care analyzer, is functioning correctly and is ready to accurately measure patient samples.
● To validate test results:
This helps confirm the reliability of results for individual patients, including their total cholesterol, HDL, LDL, and triglyceride levels.
Test Method
Allow all test materials to reach an operating temperature of 20- 40°C prior to testing. Refer to the BioAid® Lipid Testing System User's Manual for detailed instructions.
● Turn on the meter.
● Insert the code chip into the meter. Refer to Coding the Meter in the User's Manual for details. Make sure the control solution is tightly closed before use.
● Check that the specimen type shown on the meter is set to blood (BL).
● Compare the code number on the code chip with the code number printed on the test device canister label or pouch label and ensure the two numbers are identical to avoid inaccurate results.
● Wait for the meter to flash the test device symbol. Insert a test device completely into the device channel in the same direction as the arrows printed on the device until it cannot be inserted any further.
● When the meter is flashing the blood drop symbol, open the screw cap of the control solution bottle and turn the bottle upside down. Squeeze the control solution bottle gently and discard the first drop. If there are bubbles in the previous drop, squeeze the bottle and discard another drop until there are no bubbles in the drop. Apply the next drop to the specimen well on the test device while keeping the bottle vertically upside down. Make sure the control solution is applied directly into the specimen well and that there is no bubble in the solution drop.
Note:
● Make sure the bottle is completely vertical when applying the solution to the device. The volume will be inconsistent if the bottle is not completely vertical.
● Gently squeeze so that the solution makes a complete drop on the tip of the bottle and falls freely into the specimen well. Avoid touching the device with the tip of the bottle to finish an incomplete drop.
● The control solution bottle need be replaced in the pouch and sealed well after each use to avoid the evaporation.
● For the 3-in-1 test, two kinds of control solutions need to be tested on two separate test devices. Remember to switch to a new test device after the control solution has been tested on the first device.
● The results will appear on the screen within 2 minutes. Refer to the User's Manual for detailed test procedures
● Make sure to choose the correct specimen type setting to continue with lipid testing.
Video
Catalog
|
Product Name |
Catalog Number |
Components |
Quantity |
|
Lipid Control Solution (for 3 in 1 test) |
LA-131-1 |
TC Control Solution 1 2mL/bottle |
1 |
|
TC Control Solution 2 2mL/bottle |
1 |
||
|
HDL/TG Control Solution 1 2mL/bottle |
1 |
||
|
HDL/TG Control Solution 2 2mL/bottle |
1 |
||
|
Control Solution Insert |
1 |
||
|
Lipid Control Solution (for TC single test) |
LA-131-2 |
TC Control Solution 1 2mL/bottle |
2 |
|
TC Control Solution 2 2mL/bottle |
2 |
||
|
Control Solution Insert |
1 |
||
|
Lipid Control Solution (for HDL/TG single test) |
LA-131-3 |
HDL/TG Control Solution 1 2mL/bottle |
2 |
|
HDL/TG Control Solution 2 2mL/bottle |
2 |
||
|
Control Solution Insert |
1 |